About this manual

This guide provides detailed instructions for configuring Thermo NanoLC systems for use with mass spectrometry. It includes schematics for various column setups, such as EASY-Spray, Liquid Junction, and Packed Emitter configurations. The guide also lists consumable part numbers and maintenance tips, including when to replace unions and sleeves. Pressure limits for RSLCnano systems are explained with settings for flow and pressure control.

What's inside

  • EASY-nLC 1000 with Precolumn and EASY-Spray Column
  • EASY-nLC 1000 with EASY-Spray Column
  • EASY-nLC 1000 with Liquid Junction, Precolumn, Analytical Column and Emitter
  • EASY-nLC 1000 with Liquid Junction, Precolumn and Packed Emitter
  • Limiting the Pressure on RSLCnano systems
  • How Frequently Should I Replace My Unions and Sleeves?

Frequently asked questions

How frequently should I replace my unions and sleeves?

Red peek unions for 1/32” OD (IDEX) P-771 and L-Tight Union (IDEX) P-720 should be replaced no more than two connections. Sleeves for connecting OD 280 um Black and OD 360 um Beige capillary to unions should be replaced every time you make a new connection. Viper Unions (6041.5290 and 6041.5289) should be replaced whenever damaged, clogged, or short.

What are the pressure limits for RSLCnano systems?

The RSLCnano has an upper and lower pressure limit that can be set in each instrument method. The lower limit can be set from 0 to 1000 bar, and the upper limit can be set from 0 to 1000 bar.