Anderson-Negele PPR Application Guide for PMO Thermometer Specifications

This document introduces how PPR is used in PMO applications, including temperature-recording devices, recorder/controllers, temperature-recording devices used in storage tanks, and temperature-recording devices on cleaning systems.


Anderson-Negele PPR Application Guide for PMO Thermometer Specifications - cover page
Brand
ANDERSON-NEGELE
Category
Other
Document type
Application Note
Language
English
Pages
17
File format
PDF
File size
930 KB
Published
19 October, 2023
Updated
19 October, 2023
MD5 checksum
705226C4EA2309A99C80BA840208F80A

About this manual

The Anderson-Negele Paperless Process Recorder (PPR) combined with Legendary software is designed to satisfy 2019 PMO requirements for temperature recording applications. The document specifically addresses the use of the PPR with the Anderson-Negele CT8V temperature sensor for batch (vat) pasteurizer applications, ensuring compliance with various regulatory standards.

The guide covers configurations for utilizing temperatures both less than and greater than 71°C (160°F). It provides specific examples of recorder setup configurations, including time periods, axis divisions, and input range values, to demonstrate how the system can meet chart scale and time accuracy requirements mandated by the PMO.

What's inside

  • TEMPERATURE-RECORDING DEVICES FOR BATCH PASTEURIZERS
  • RECORDER/CONTROLLERS FOR CONTINUOUS PASTEURIZERS
  • TEMPERATURE-RECORDING DEVICES USED IN STORAGE TANKS
  • TEMPERATURE-RECORDING DEVICES ON CLEANING SYSTEMS
  • UTILIZING TEMPERATURES LESS THAN 71°C
  • UTILIZING TEMPERATURES GREATER THAN 71ºC
  • Temperature Accuracy Requirements
  • Time Accuracy Requirements
  • Self-diagnostic circuitry
  • Tamper Evident Sealing Covers

Specifications

Temperature AccuracyWithin ± 0.5°C (± 1° F), between 60°C (140°F) and 69°C (155°F)
Chart Scale SpanNot less than eleven (11) Celsius degrees (twenty (20) Fahrenheit degrees)
Chart Scale Divisions0.5°C (1°F), spaced not less than 1.6 millimeter (0.0625 of an inch) apart
NEMA RatingNEMA 4X rated
Chart Revolution TimeNot more than twelve (12) hours

Frequently asked questions

Can the PPR be used as an indicating thermometer in PMO applications?

No, the PPR is not intended to be used as an indicating thermometer in PMO applications.

Which temperature sensor is recommended for the PPR to meet PMO requirements?

The PPR can be used with an Anderson-Negele CT8V temperature sensor to meet the listed requirements.

How does the PPR protect calibration against unauthorized changes?

The PPR is equipped with a movable shunt jumper 'J9' to toggle between RUN and PROGRAM modes, inhibiting calibration in RUN mode.