Anderson-Negele ITM-51-CU Turbidity Sensor Calibration and Maintenance Manual

This document is the manual of Anderson-Negele ITM-51-CU turbidity sensor, which introduces the application, usage, maintenance and other content of the sensor


Anderson-Negele ITM-51-CU Turbidity Sensor Calibration and Maintenance Manual - cover page
Brand
ANDERSON-NEGELE
Document type
Service Manual
Language
Chinese
Pages
2
File format
PDF
File size
205 KB
Published
19 October, 2023
Updated
19 October, 2023
MD5 checksum
A6071573C57BDA73A3A9CBF19B2BDC36

About this manual

This manual provides instructions for calibrating and maintaining the ITM-51-CU turbidity sensor, including steps for checking calibration with the solid-state reference plate. It also details the scope of delivery, including the reference plate and cap components.

The document recommends annual calibration checks and outlines procedures for cleaning the lens, setting measurement parameters, and verifying displayed values against reference labels. If calibration drifts beyond ±6%, the sensor should be returned for recalibration. This manual is written in Chinese.

What's inside

  • Application/intended use
  • Scope of delivery
  • Calibration check
  • Order structure and spare parts

Frequently asked questions

How often should the ITM-51-CU turbidity sensor be calibrated?

It is recommended to check the calibration once a year.

What tool is recommended for calibration checks?

The ITM-51-CU calibration check tool is recommended for calibration checks.

What should be done if the displayed NTU value is outside the limits?

It is recommended to send back the ITM-51 for recalibration.