This manual introduces the use instructions and operation flow of the DSD®-201 and DSD-201LT MEDIVATORS endoscope reprocessing system, including device functions, operating specifications, daily maintenance and other content.
This document serves as an in-service guide for the Medivators DSD-201 and DSD-201LT endoscope reprocessing systems, intended for facility use only. It outlines the intended use for high-level disinfection of flexible immersible endoscopes and details the necessary safety precautions and guidelines for reliable disinfection.
| Intended Use | high level disinfection of flexible immersible endoscopes |
|---|---|
| Water Pressure Requirement | minimum of 40 psi |
| Water Pressure Range (Flush) | 35-40 psi |
| Water Pressure Range (Regulator) | 40-45 psi |
| HLD Volume Required | 4 gallons |
| Water Filter Replacement Interval | every 3 months or if below 40 psi |
| Air Filter Replacement Interval | Six Months |
| 0.2 micron Water Filter Replacement | every 6 months or if below 40 psi |
The system is intended for the high-level disinfection of flexible immersible endoscopes.
The 1 micron and 0.45 micron water filters should be replaced every 3 months or if the pressure drops below 40 psi.