Medivators DSD-201 Manuals

ManualLib has 2 documents for the Medivators DSD-201 (medical technology): 2 user manuals. Every file can be read online or downloaded as a PDF for free.

User Manual 2

Key facts

Model
DSD-201
Documents
2
Languages
ZH · EN
Last added
2026-10-29

Start here

Medivators DSD-201 Endoscope Washer Disinfector User Manual

Medivators DSD-201 Endoscope Washer Disinfector User Manual

User Manual 98 pages · 3 MB · ZH

The document serves as a comprehensive guide for the Medivators DSD-201 Endoscope Washer Disinfector. It details the machine's features, setup procedures, daily operation cycles (including cleaning, disinfection, rinsing, and drying phases), and safety guidelines for operators and maintenance personnel.

Key sections cover the operator controller interface, including LCD display functions and digital keypad settings. The manual provides step-by-step instructions for preparing endoscopes, running disinfection programs, monitoring chemical efficacy, and performing necessary maintenance tasks such as filter replacement and air filter cleaning. This manual is written in Chinese.

Read online

Medivators DSD-201 specifications

Dimensions (H x W x D)46"x36"x21" (117x91x53 cm)
WeightApprox. 400 lbs (181 kg)
Power Supply230VAC 50/60Hz, 6amp or 120VAC 60Hz, 12 amp
Water Quality RequirementMunicipal drinking water (cold supply)
Disinfectant Container Capacity15 liters

Source: Medivators DSD-201 Endoscope Washer Disinfector User Manual

All Medivators DSD-201 documents by type

User Manual 2

Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.

Questions answered in these manuals

What is the minimum water quality requirement for this machine?

The machine requires municipal drinking water (cold supply) as a minimum standard.

From: Medivators DSD-201 Endoscope Washer Disinfector User Manual

How often should the 0.2-micron filter be replaced?

It is recommended to replace the 0.2-micron filter every 6 months or sooner, depending on water quality.

From: Medivators DSD-201 Endoscope Washer Disinfector User Manual

What is the intended use of the DSD-201 and DSD-201LT?

The system is intended for the high-level disinfection of flexible immersible endoscopes.

From: Medivators DSD-201 & DSD-201LT Endoscope Reprocessing System User Manual

How often should the water filters be replaced?

The 1 micron and 0.45 micron water filters should be replaced every 3 months or if the pressure drops below 40 psi.

From: Medivators DSD-201 & DSD-201LT Endoscope Reprocessing System User Manual