User manual for the Medivators CER endoscope reprocessor, covering installation, operation, safety, maintenance and specifications.
The Medivators CER is a tabletop automatic endoscope reprocessor designed for cleaning, disinfection and rinsing of flexible endoscopes. This manual provides detailed instructions for safe operation, proper installation, water and chemical handling, and routine maintenance to ensure reliable high‑level disinfection.
Key sections include safety guidelines, step‑by‑step operating procedures, electrical and piping requirements, water‑filter specifications, chemical usage, and troubleshooting. The document also contains a full list of specifications, warranty information and a glossary of terms. This manual is written in Chinese.
| Dimensions (closed) | 16 x 22 x 20 in |
|---|---|
| Dimensions (open) | 33 in (height) |
| Weight (120 V) | 77 lb (35 kg) |
| Weight (230 V) | 81 lb (37 kg) |
| Electrical requirement | A.C. 230‑240 V, 50/60 Hz, 840 VA |
| Water consumption with OPA disinfectant | 20 gal (75 L) for full cycle + 2 rinses |
| Minimum water pressure | 40 psi (2.8 bar) |
The 0.2 µm water filter should be replaced every 6 months or sooner, depending on the pre‑filter system and inlet water quality.
Before moving the CER, disconnect the power cord, chemical lines, drainage and water supply lines, and keep the unit upright to avoid damage.
The supply water pressure must be at least 40 psi (2.8 bar).
Operators should wear appropriate personal protective equipment to avoid biological contamination and chemical burns.