This guide introduces the techniques of using multi-angle static light scattering for size exclusion chromatography (SEC-MALS) measurement of macromolecular mass in solution.
This guide explains how to prepare samples, set up the SEC-MALS system, and collect data for accurate macromolecular mass measurements. It covers instrument components, buffer requirements, column compatibility, and startup procedures. The system combines size exclusion chromatography with multi-angle light scattering and concentration detectors for precise mass and oligomeric state analysis.
Key features include a Wyatt Dawn Heleos II MALS detector, RI and UV detectors, and an Agilent chromatography system. Users must prepare samples in compatible buffers, filter them, and follow startup steps including column equilibration and detector calibration.
| MALS Detector | Wyatt Dawn Heleos II |
|---|---|
| Refractive Index Detector | Wyatt Optilab T-rEX |
| Chromatography System | Agilent 1260 Infinity Isocratic Liquid Chromatography System |
| UV Detector | Agilent variable wavelength UV detector |
| Column Type | Silica or agarose columns |
| Maximum Injection Volume | 100 µl |
| Minimum Injection Volume | 5 µl |
| Typical Protein Concentration Range | 5 – 500 µg/injection |
It measures macromolecular mass in solution by combining size exclusion chromatography with multi-angle static light scattering.
PBS or HBS (25 mM HEPES pH 7-7.5, 150 mM NaCl), filtered.