Pall Allegro Single-Use Mixer Validation Guide USTR 2777a

USTR 2777a Validation Guide Pall Allegro Single-Use Mixers introduces the validation overview of Single-Use Mixers, including water leak test, drainage and drain valve operation, resistance to gamma sterilization, connections testing, extractables study, biological safety tests, physico-chemical tests and shelf life studies


Pall Allegro Single-Use Mixer Validation Guide USTR 2777a - cover page
Brand
PALL
Document type
Application Note
Language
English
Pages
28
File format
PDF
File size
707 KB
Published
19 October, 2023
Updated
19 October, 2023
MD5 checksum
5ECBE8F1D675D7278ECE7190C18C25A2

About this manual

This validation guide presents the testing performed on Pall Allegro single‑use mixers (200 L, 500 L and 1000 L) to demonstrate suitability for biotechnology and pharmaceutical use. It includes water leak tests, drainage performance, resistance to gamma sterilization, connections testing, extractables studies, biological safety, physico‑chemical tests and shelf‑life studies.

The document details test methods, results and conclusions, confirming that the mixers meet USP standards, can operate continuously for three days, and tolerate gamma doses up to 55 kGy. Updated surface‑area data for the upgraded 500 L and 1000 L models are also provided.

What's inside

  • Validation Overview
  • Water Leak Test
  • Drainage and Residual Drainage Testing
  • Resistance to Gamma Sterilization
  • Connections Testing
  • Extractables Testing
  • Biological Safety Tests
  • Physico‑chemical Tests
  • Shelf Life Studies
  • Appendix A