This is a calibration manual for the HBC-4301A Hot Bonding Controller
This manual provides detailed instructions for calibrating the Novatech Controls HBC-4301A Hot Bonding Controller. It covers calibration of ADCs, thermocouple types, cold junction sensors, vacuum and pressure sensors, and power output levels. The procedure requires specific test equipment and must be performed by qualified personnel. Safety checks, including earth leak detectors and socket verification, are also included. Calibration values are stored permanently in the device.
| Reference Voltage #1 | 53.6mV +/- 2.0 |
|---|---|
| Reference Voltage #2 | 481mV +/- 10 |
| Thermocouple Type | K-Type or J-Type |
| Accuracy | +/- 1°C |
The manufacturer recommends calibration every 12 months.
Required equipment includes an RTD sensor Class A, thermocouple output generator, digital multimeter, 4-20mA loop powered device, vacuum measuring instrument, and pressurised air supply.
Press and hold the Setup Key, then turn on the power. Hold until 'Calibration Menu Enabled' appears. Release the key, wait for initialization, then press and hold Setup Key again until 'Entering Calibration Menu' appears.