Greenspan 7SDI-1000 SDI Converter Operation and Configuration Manual

The 7SDI-1000 is a Greenspan sensor converter that can convert the serial RS232 interface to an SDI protocol electrical interface and command protocol. The converter connects to the sensor via an HS7 connector and provides power to the sensor through that connector. It is possible for the converter to turn sensor power off using an SDI12 command.


Greenspan 7SDI-1000 SDI Converter Operation and Configuration Manual - cover page
Brand
Greenspan
Category
Industrial Robot
Document type
User Manual
Language
English
Pages
13
File format
PDF
File size
1010 KB
Published
18 October, 2023
Updated
18 October, 2023
MD5 checksum
28170A77E4C1E73CE53E7AA2E895879F

About this manual

This manual explains how to operate and configure the Greenspan 7SDI-1000 SDI-12 converter, which interfaces with Greenspan sensors via an HS7 connector. It supports SDI-12 protocol commands and allows power control, sensor identification, and firmware updates. The converter works with specific Greenspan sensor models like PS3100, CTD3100, and PS7000. Setup involves connecting sensors via RS232 or bare wire, configuring sensor models using SDI-12 commands, and setting data output order and precision.

The manual includes detailed command syntax for configuring the converter, such as setting sensor model, data channel order, and power control. It also covers wiring connections using the 5CC-770 adaptor cable for bare wire sensors and provides examples of SDI-12 commands for data retrieval and sensor status checks.

What's inside

  • SDI-12 Protocol Support
  • Sensor Compatibility
  • Power Control
  • Wiring and Connections
  • Command Syntax
  • Sensor Setup
  • Data Channel Configuration
  • Firmware Update

Specifications

Supported ProtocolSDI-12 Version 1.3
Operating Voltage9 to 30V DC
Sensor Models SupportedGreenspan 3000 Series Sensors (PS3100, CTD3100, TS3000, EC3000, ODO3000, PH3000, ORP3000, DO3000, MP3000) and Other Greenspan Pressure Sensors (PS2100, PS1000, PS7000)

Frequently asked questions

What sensors is the 7SDI-1000 compatible with?

It is compatible with Greenspan 3000 Series Sensors (PS3100, CTD3100, TS3000, EC3000, ODO3000, PH3000, ORP3000, DO3000, MP3000) and other Greenspan pressure sensors (PS2100, PS1000, PS7000).

How do I set up the 7SDI-1000 for a PS7000 sensor?

Use the command 0XM1! to set the model number to PS7000. Then verify with the 0I! command to confirm the sensor model is set correctly.

What is the power range for the 7SDI-1000?

The operating voltage range is typically 9 to 30V DC.