This manual describes the functions and usage of Greenspan pH sensors
This manual provides instructions for using the Greenspan PH100 pH sensor, including calibration procedures, connection diagrams, and housing maintenance. The sensor uses a gel-filled electrode and is temperature compensated from 0°C to 50°C. Calibration requires disassembly to adjust potentiometers for zero and span settings using pH4, pH7, or pH10 solutions. The sensor is designed for field use with a 12V power supply and includes protection against reverse polarity and lightning surges.
The manual also covers housing disassembly and reassembly, electrode replacement, and packaging details. It warns against dismantling the sensor without authorization and outlines warranty conditions and limitations, including exclusions for damaged or altered components.
| Output Range | 4-20mA |
|---|---|
| Temperature Compensation | 0°C to 50°C |
| Supply Voltage | 8-15V |
| Current Draw | 45mA quiescent + 20mA full scale |
| Cable Resistance | 9 ohms per 100 metres per conductor |
The output current for pH 7 is 12.00mA.
The sensor is temperature compensated from 0°C to 50°C.