This document introduces the installation instructions for the MK4 and IK4/SK4/RK4/RK2 series transducers, including installation precautions, correct use of installation brackets, correct use of floating cursors, installation of transducers, etc.
This manual provides installation instructions for Gefran MK4, IK4, SK4, RK4, and RK2 series magnetostrictive transducers. It covers general precautions, the correct use of installation brackets and floating cursors, and specific mounting procedures for rod-type transducers inside hydraulic cylinders.
| Refresh rate | up to 1000 measurements/second |
|---|---|
| Output resolution | 24-25 bit |
| Max torque for fastening screws | 1.1 Nm |
| Minimum distance between multiple cursors | 75 mm |
| Recommended spacer thickness for magnetic support | 15 mm |
The cursor must be installed on a support made of non-magnetic material such as brass, aluminium, or AISI316 stainless steel.
The fastening screws must be tightened with a maximum torque of 1.1 Nm.
The SSI output goes to 0 if the echo is absent, which indicates the magnet is out of the measurement range or there is an internal device error.