This is the operating manual of Medi-Therm II hyper/hypothermia machine
The Medi-Therm II Hyper/Hypothermia machine regulates patient temperature by supplying temperature-controlled water through a connector hose to a Gaymar Hyper/Hypothermia blanket. The system includes a patient probe to sense temperature and an operator control panel for managing therapy settings.
| Maximum blanket temperature (MTA5900) | 41°C (105.8°F) |
|---|---|
| Maximum blanket temperature (MTA5942) | 42°C (107.6°F) |
| Minimum blanket temperature | 4°C (39.2°F) |
| Low probe temperature alert threshold | below 32°C [89.6°F] |
| High probe temperature alert threshold | above 45°C [113°F] |
Only distilled water should be used to fill the machine.
The attending physician must be notified promptly.
The YSI 400 series probe or an equivalent is accepted.