About this manual

The Medi-Therm II Hyper/Hypothermia machine regulates patient temperature by supplying temperature-controlled water through a connector hose to a Gaymar Hyper/Hypothermia blanket. The system includes a patient probe to sense temperature and an operator control panel for managing therapy settings.

What's inside

  • Safety Precautions
  • Description
  • Features
  • Operator Control Panel
  • Start-up Procedure
  • Shutdown Procedure
  • Modes of Operations
  • Patient Safety
  • Care of Medi-Therm II
  • Alerts/Troubleshooting

Specifications

Maximum blanket temperature (MTA5900)41°C (105.8°F)
Maximum blanket temperature (MTA5942)42°C (107.6°F)
Minimum blanket temperature4°C (39.2°F)
Low probe temperature alert thresholdbelow 32°C [89.6°F]
High probe temperature alert thresholdabove 45°C [113°F]

Frequently asked questions

What type of water should be used to fill the machine?

Only distilled water should be used to fill the machine.

What happens if the patient's temperature does not reach the prescribed level?

The attending physician must be notified promptly.

Which patient probe is compatible with the Medi-Therm II?

The YSI 400 series probe or an equivalent is accepted.