This document introduces the use instructions and precautions of Medi-Temp II FW400/FW401/REF FW403 (FW400 Series) Blood/Fluid Warmer.
The Gaymar FW400 Series Blood/Fluid Warmer is a sealed, fluid-resistant device designed for the safe and rapid in-line warming of blood, blood products, and other fluids. It operates using disposable Blood/Fluid Warming Sets and features an LED temperature display and a lighted on/off circuit breaker. The device is intended to prevent inadvertent hypothermia during fluid administration in medical settings.
| Operating Temperature Range | 36°C to 43°C |
|---|---|
| Minimum Input Temperature | 10ºC |
| Maximum Flow Rate (Standard) | 300 ml/min |
| Maximum Flow Rate (Room Temp) | 500 ml/min |
| Priming Volume (Standard) | 39 ml |
| Priming Volume (High Flow) | 79 ml |
| Warranty Period | 1 year |
This device is intended to aid in the prevention of inadvertent hypothermia during administration of blood, blood products, and other fluids.
No, do not immerse the unit in cleaning or sterilization solution as it is fluid-resistant, not fluid-proof.
The warmer may not be heating; refer to the Safety Systems section for troubleshooting.
No, the bubble trap must be mounted and kept in a vertical position at all times to prevent air introduction to the patient.