This is a manual supplement document containing revisions for the 726 user manual.
This document is a manual supplement for the Fluke 726 calibrator, designed to ensure the accuracy of the main user manual. It contains specific instructions for updating information regarding display indicators, configuration settings such as step and ramp times, and operational procedures for shut-down modes and HART resistor toggling.
| Pulse Frequency Max | 15 kHz |
|---|---|
| Pulse Frequency Range | 2 CPM to 15 kHz |
| Step Time Range | 1 sec to 99 sec |
| Ramp Time Range | 5 sec to 99 sec |
| EMC Standard | IEC 61326-1 |
The maximum frequency for the Pulse function is 15 kHz.
Press the menu button until STEP TIME appears, use the arrow keys to set the time between 1 and 99 seconds, and press save.