This is a manual for αlpha CON 190 controller/transmitter.
This manual provides step-by-step instructions for operating and calibrating the αlpha CON 190 1/8 DIN Conductivity Controller. It includes sections on safety, front and back panel components, wiring, measurement modes, setup functions, relays, and specifications. The controller features two set points, automatic/manual temperature compensation, and isolated 4-20mA output. It is designed for panel mounting and supports conductivity ranges from 0.00 to 2000 µS/cm and temperature from -10 to 110°C.
| Conductivity Range | 0.00 – 20.00 µS/cm to 0.0 – 200.0 mS/cm |
|---|---|
| Temperature Range | -10 to 110 °C |
| Temperature Resolution/Accuracy | 0.1 / ± 0.5 °C |
| Power Requirements | 85 to 260 VAC, 50/60 Hz or DC |
| Relays | 2 SPDT relays |
| Signal Output | 4-20 mA galvanically isolated |
| Dimensions | 96 x 48 x 150 mm |
The conductivity range is from 0.00 to 20.00 µS/cm to 0.0 to 200.0 mS/cm, with selectable ranges and resolutions.
The temperature range is -10 to 110°C with a resolution of 0.1°C and accuracy of ±0.5°C.
The controller accepts 85 to 260 VAC, 50/60 Hz or DC.
The controller has two SPDT relays rated for 6 amps at 110 VAC.