This manual introduces the usage of the DO 6+ hand held meter, including usage steps and reference guide.
This manual provides step-by-step instructions for operating the Eutech DO 6+ handheld meter, including setup, calibration, measurement, and probe maintenance. It covers both percentage saturation and concentration (mg/L) modes, with temperature compensation and salinity/pressure adjustments. The device is designed for field use with a splash-proof keypad and large LCD display.
The manual also includes troubleshooting, error messages, factory defaults, specifications, and warranty information. It is intended for users who need to measure dissolved oxygen in water applications such as aquaculture, wastewater treatment, and boiler water monitoring.
| % Saturation Mode Range | 0.00 – 200.0 % |
|---|---|
| Concentration Mode Range | 0.00 – 20.00 mg/L or ppm |
| Temperature Range (Meter Only) | -5.0 – 105.0 °C |
| Salinity Correction Range | 0.0 – 50.0 ppt |
| Barometric Pressure Correction Range | 500 to 1499 mm Hg or 66.6 to 199.9 kPA |
| Probe Type | Galvanic / Thermistor |
| Response Time | 60 seconds to achieve 95% of the reading |
The DO 6+ offers measurement of dissolved oxygen as percentage saturation or concentration (mg/L or ppm), and temperature in degrees Celsius.
Calibrate using 100% saturated air or air-saturated water for the 100% point and 0% oxygen solution for the 0% point.
The meter comes with a three-year warranty, and the electrode has a six-month warranty against significant deviations in material and workmanship.