Elma 291 User Manual
This manual provides step-by-step instructions for operating the Elma 291 electrostatic field meter, including grounding, measurement procedures, and alarm settings. It covers safety precautions, instrument description, and maintenance tasks such as battery replacement and disposal. The device measures static voltage from 0.010kV to 20,000kV with a sensor distance of 2.54cm.
| Display | LCD with 4 1/2 digit display up to 19999 |
|---|---|
| Resolution | 0.001kV |
| Measurement Range | -19.999kV ~ +19.999kV |
| Accuracy | ±(5% + 10dgt) |
| Overload Indication | OL |
| Test Interval | 2 times per second |
| Battery Type | 9V NEDA 1604, IEC 6F22 or JIS 006P |
| Battery Life | Approx. 100 hours |
| Operating Temperature and Humidity | 5 to 40°C (41 to 104°F); < 80%RH |
| Storage Temperature and Humidity | -10°C to 60°C (14 to 140°F); < 70%RH |
| Weight | Approx. 210g |
| Dimensions | 148(L) x 65(W) x 37.5(H)mm |
Use one of two methods: connect the grounding wire to a building’s earth or electrical socket, or use an anti-static wrist strap connected to earth.
The measurement range is 0.010kV to 20,000kV with a sensor distance of 2.54cm (1 inch).
Navigate to SET 1 for high alarm threshold, SET 2 for low alarm threshold, and SET 3 to enable or disable the alarm.
Replace the battery immediately to ensure accuracy.