This is a user manual that introduces the functions, operations, precautions and other information of ELMA COMBITEST 422.
This manual provides instructions for the ELMA COMBITEST 422, an instrument designed for measuring and controlling the safety of electrical systems. It covers general rules, preparation, and detailed operating procedures for various tests including RPE, insulation resistance, loop impedance, and RCD testing.
| Test current for RPE | greater than 200mA |
|---|---|
| Open circuit voltage for RPE | between 4V and 24V |
| Insulation resistance test voltages | 50V, 100V, 250V, 500V or 1000V DC |
| Test current for LoΩ | greater than 10A |
| Power supply | 6x1.5V alkaline batteries AA LR06 |
| Auto shut-off time | 5 min |
Do not perform voltage or current measurements in humid environments and avoid contact with exposed metal parts.
After long storage under extreme conditions, allow the instrument to acclimatize before use.
The instrument has an auto shut-off function that can be deactivated.