This manual will assist you in installing, using and maintaining the Eletta flow meter. It is your responsibility to make sure that all operators have access to adequate instructions about safe operating
This manual provides instructions for the installation, operation, and maintenance of the Eletta EMFA2E21 electromagnetic flow meter. The device is designed for measuring electrically conductive fluids in various industries, consisting of a detector and a converter. It is available in compact and remote versions with various power supply options.
| Measuring principle | Faraday's law |
|---|---|
| Application range | Electrically conductive fluids |
| Primary measured value | Flow velocity |
| Secondary measured value | Volume flow |
| Measurement range | -12 ... +12 m/s |
| Calibration range | 0.3-6m/s |
| Inlet run requirement | 5DN |
| Outlet run requirement | 2DN |
The meter is intended for measuring electrically conductive fluids.
An equivalent to 5 diameters is required on the inlet side and 2 diameters on the outlet side.
For best results, the meter should be calibrated at least 1 time per year.