This manual describes the Millicell® ERS-2 electrical resistance system for measuring cell resistance and cell voltage.
The Millicell ERS-2 system is a meter and electrode system designed to measure Trans Epithelial Electrical Resistance (TEER) and voltage of epithelial cells in culture. It utilizes an alternating current (AC) source to ensure accurate resistance measurements without affecting cell membrane integrity or causing electrochemical deposition. The system consists of a meter with an LCD display and analog output, and interchangeable electrodes (STX01, STX04, STX00) for different well formats.
| Battery Type | 6 V NiMH 2,700 mAh |
|---|---|
| Battery Charger Voltage | 12 V DC |
| Battery Charge Duration | 8–10 hours |
| Max Resistance Reading | 13,000 Ω |
| Guaranteed Accuracy Range | up to 9,999 Ω |
| Voltage Offset Limit | 20 mV |
The battery should be charged overnight before intended use.
No, the electrode cannot withstand high temperatures and should not be autoclaved or flame sterilized.
The electrode should be soaked in solution for 24 hours if it has been stored dry.
The electrode should be re-equilibrated for 12–24 hours while attached to the meter with the power off.