Envirologix AQ-170-BG Corn Bulk Grain HS Enogen QuickGuide

This document introduces the test method for AQ-170-BG Corn Bulk Grain HS Enogen product.


Envirologix AQ-170-BG Corn Bulk Grain HS Enogen QuickGuide - cover page
Brand
Envirologix
Document type
User Manual
Language
English
Pages
1
File format
PDF
File size
1 MB
Published
17 October, 2023
Updated
17 October, 2023
MD5 checksum
86AF49095969F8814E844B55E22B3CAC

About this manual

The document outlines a specific test procedure for detecting Enogen in corn bulk grain samples. It details the requirements for sample collection, grinding, and sub-sampling to ensure accuracy, specifying that at least 95% of the ground sample must pass through a 20 mesh sieve.

The guide describes the step-by-step process for preparing the test extract using EB17 pouches and distilled water. It includes specific instructions on shaking, centrifugation (at 2000 x g), incubation at 22°C, and the final reading steps using a QuickScan device to determine if the result is below or above 0.04%.

What's inside

  • Test Procedure
  • Sample Preparation
  • Extract Preparation
  • Incubation
  • QuickScan Reading

Frequently asked questions

What is the required sample preparation for corn grain testing?

You must collect and grind a representative corn sample so that at least 95% of it passes through a 20 mesh sieve.

How much water and how many dissolvable pouches are needed for the test?

Add one dissolvable pouch of EB17 and 75 mL of distilled, deionized, or bottled water to the container.

What are the incubation temperature requirements for the test?

The incubator must reach 22°C, and acclimation is required if the ambient temperature is unknown or outside the 20-24°C range.

What are the possible results returned by the QuickScan?

The QuickScan will return a result that is either less than 0.04% or greater than 0.04%.