This is an instruction manual about how to prepare samples and test them using the EnviroLogix Fumonisin Flex DDGS kit
This document provides a sample preparation test procedure for the EnviroLogix Fumonisin Flex DDGS kit. It details the required accessories, such as 1 mL and 50 µL pipettes, and specifies the use of EB18 and DB6 Buffer reagents. The procedure covers the extraction of a sub-sample, mechanical shaking, and centrifugation steps to prepare the sample for analysis.
| Base Range | 1.5-7 ppm |
|---|---|
| Centrifuge Speed | 2000 x g |
| Acclimation Temperature | 20-24°C (68-75°F) |
The base range results are valid between 1.5 and 7 ppm.
Base range results are considered valid between 1.5 and 7 ppm.
You must remove the strip from the vial immediately after 5 minutes.
Acclimation is required when ambient temperature is unknown or outside of 20-24°C (68-75°F).