RAE Systems SP-2104 Toxic Gas Detector Operation and Maintenance Manual

The RAE Systems SP-2104 is a fixed electrochemical toxic gas detector with 4‑20 mA, RS‑485 outputs and LED display, suitable for industrial environments.


RAE Systems SP-2104 Toxic Gas Detector Operation and Maintenance Manual - cover page
Brand
RAE Systems
Category
Other
Document type
User Manual
Language
Chinese
Pages
2
File format
PDF
File size
118 KB
Published
30 September, 2026
Updated
30 September, 2026
MD5 checksum
3E0916A1AFB2E3A14C9827CC3C672BA6

About this manual

The SP-2104 is a fixed‑type toxic gas detector that uses an electrochemical three‑electrode sensor to monitor gases such as carbon monoxide, hydrogen sulfide, sulfur dioxide and hydrogen. It features an explosion‑proof aluminum alloy housing, IP65 protection and a bright LED display for on‑site monitoring.

Key specifications include a 4‑20 mA current loop, RS‑485 digital output and a 30 V 2 A relay, powered by a DC 10‑30 V supply with less than 2 W consumption. The device operates from –20 °C to +50 °C, tolerates 10‑93 % humidity, and meets Ex d IICT6 CMC explosion‑proof certification. This manual is written in Chinese.

What's inside

  • Detection Principle
  • Detection Range
  • Electrical Interface
  • Power Supply
  • Output Options
  • Physical Dimensions
  • Environmental Ratings
  • Standard Configuration
  • Technical Parameters

Specifications

Detection principleElectrochemical
CO detection range0–1000 ppm
H2S detection range0–200 ppm
SO₂ detection range0–100 ppm
H₂ detection range0–1000 ppm
Standard output4‑20 mA
Digital outputRS‑485
Relay output30 V 2 A
Power supply voltageDC 10‑30 V
Maximum load resistance260 Ω
LED display resolution0.01 (range 50‑1000)
Protection ratingIP65
Explosion protectionEx d IICT6 CMC
Enclosure materialAluminum alloy
Dimensions (H×W×D)239 × 185 × 118 mm
Weight2200 g
Operating temperature‑20 °C to +50 °C
Relative humidity10 %–93 %
Atmospheric pressure86 kPa–106 kPa

Frequently asked questions

What is the required power supply voltage?

It operates from a DC supply of 10 V to 30 V.

What protection rating does the detector have?

The enclosure is rated IP65 and meets Ex d IICT6 CMC explosion‑proof certification.

Download PDF (118 KB)

More RAE Systems Manuals