This document is a maintenance manual for the Harvard Electronic balance.
This document serves as an Operator's, Unit, and Direct Support Maintenance Manual for the Ohaus Harvard Trip Balance. It details the setup process, including leveling and zeroing the scale, and explains the weighing procedure using the beam poises and optional tare poise. The manual also covers specific gravity determination using an accessory clamp and rod support.
| Warranty Period | ninety (90) days |
|---|---|
| Bearing Material | polished agate V-blocks |
| Knife Edge Material | hardened, precision-ground steel |
It is advisable to check the zero balance periodically because foreign material may accumulate on the platforms or the beams.
Remove any mass or weights from the platform(s) and move one of the poises away from zero to stop oscillation.
No, the bearings should never be lubricated as they are designed to work most efficiently when clean and dry.