Introduces Pioneer™ Balance safety precautions and installation steps
The document serves as an instruction manual for Pioneer electronic balances. It covers safety precautions, control functions, and detailed steps for installing components such as the pan support and glass doors depending on the model capacity.
The guide explains how to level the balance and connect power, followed by procedures for initial calibration using internal or external masses. It details various weighing modes including Count, Percent, and Weigh Below features, along with extensive settings for menu navigation, unit selection, RS232 communication parameters, and security locking.
| Input Voltage Requirement | Verify that the AC Adapter input voltage matches the local AC power supply. |
|---|---|
| Operating Environment | Use only in dry locations and avoid hostile environments. |
| Warm-up Time for Calibration | 60 minutes |
| RS232 Baud Rates | 600, 1200, 2400, 4800, 9600 and 19200 |
InCal or Span calibration should be performed daily and when the room temperature changes.
Press and hold the No button at any time to exit quickly.
It assumes the initial item placed on the balance is a container so it is zeroed out before weighing the next item.