Fluke 1550C/1555 Insulation Tester Calibration Manual
This document serves as a calibration manual for the Fluke 1550C/1555 Insulation Tester. It outlines the necessary safety precautions, required equipment, and step-by-step performance test procedures to verify the device's operating condition. The manual details how to test the IR port, buttons, display, charging, insulation accuracy, output voltage, short circuit current, and voltage measurement accuracy.
| Warranty Period | three years |
|---|---|
| Software Warranty | 90 days |
| Parts Warranty | 90 days |
| HV Probe Requirement | 6 kV, ±1 % (1000:1 Divider) |
| HV Probe Requirement (1555) | 11 kV, ±2 % |
| Digital Multimeter Requirement | 500 mVdc to 1 V: ±0.02 % |
| Calibrator DC Voltage | 0 - 550 V |
| Calibrator AC Voltage | 0 - 240 V, 60 Hz |
The warranty period is three years and begins on the date of shipment.
No, do not attempt to use the 5500A, 5520A, or other standard calibrator for insulation and continuity resistance tests as it will cause damage.