This manual introduces how to use the 719Pro pressure calibrator for calibration
The document serves as a comprehensive guide for technicians to verify and calibrate the Fluke 719Pro Pressure Calibrator. It details specific verification procedures for pressure, DC voltage, DC current, RTD measurements, and mA loop performance using required test equipment.
Additionally, the manual covers maintenance tasks such as battery replacement and cleaning of the product and valve assembly. It includes specifications for various pressure ranges (30G, 150G, 300G) along with accuracy ratings and temperature effects.
| Maximum Pressure Range | 300 PSI / 20 bar |
|---|---|
| Pressure Accuracy (6 months) | ±0.025 % of full scale |
| mA Range | 0 mA dc to 24 mA dc |
| RTD Measurement Range | -50 q C to 150 q C |
| Battery Type | 8 AA batteries (alkaline) |
The warranty period is three years and begins on the date of shipment.
Reduce or vent the pressure immediately to prevent product damage or personnel injury.