Fluke 719Pro Pressure Calibrator Calibration Manual

This manual introduces how to use the 719Pro pressure calibrator for calibration


Fluke 719Pro Pressure Calibrator Calibration Manual - cover page
Brand
Fluke
Document type
Service Manual
Language
English
Pages
28
File format
PDF
File size
581 KB
Published
15 October, 2023
Updated
15 October, 2023
Model
719Pro
MD5 checksum
27666DF2D65FE0DF933D3C30C1CDD039

About this manual

The document serves as a comprehensive guide for technicians to verify and calibrate the Fluke 719Pro Pressure Calibrator. It details specific verification procedures for pressure, DC voltage, DC current, RTD measurements, and mA loop performance using required test equipment.

Additionally, the manual covers maintenance tasks such as battery replacement and cleaning of the product and valve assembly. It includes specifications for various pressure ranges (30G, 150G, 300G) along with accuracy ratings and temperature effects.

What's inside

  • Introduction
  • Safety Information
  • Specifications
  • Maintenance
  • Verification Procedure
  • Pressure Verification
  • DC Voltage Measure Verification
  • Calibration Adjustment

Specifications

Maximum Pressure Range300 PSI / 20 bar
Pressure Accuracy (6 months)±0.025 % of full scale
mA Range0 mA dc to 24 mA dc
RTD Measurement Range-50 q C to 150 q C
Battery Type8 AA batteries (alkaline)

Frequently asked questions

What is the warranty period for Fluke products?

The warranty period is three years and begins on the date of shipment.

How should I handle a pressure reading showing 'OL'?

Reduce or vent the pressure immediately to prevent product damage or personnel injury.

Download PDF (581 KB)

More Fluke Manuals


View all Fluke manuals →