This document introduces the use of 5500A Multi-Product Calibrator, including replacing the fuse, selecting line voltage, front panel features, rear panel features, setup menu, operate and standby, connecting the calibrator to a UUT, setting the output, waveform types, adjusting the phase, input offset, using the 5725A amplifier, editing and error output settings
The Fluke 5500A Multi-Product Calibrator is designed to operate from AC power sources up to 264V rms. It features a two-line backlit LCD for output display and a multipurpose control display for data entry and prompts. The instrument includes controls for operating modes, grounding connections, and amplifier enablement.
| AC Power Source Limit | not apply more than 264V ac rms between the supply conductors or between either supply conductor and ground. |
|---|---|
| Fuse Rating (110/120 V) | 2.5 ampere/250 volt time delay |
| Fuse Rating (220/240 V) | 1.25 ampere/250 volt time delay |
| Line Voltage Settings | 100V, 120V, 200V, and 240V (47 to 63 Hz) |
| Output Display | Two-line backlit LCD showing output amplitudes, frequency and calibrator status. |
To check or change the setting, note the voltage visible through the window in the power line fuse compartment cover and follow the procedure for Changing Line Voltage.
Pressing this key places the 5500A in the standby mode and internally disconnects the NORMAL and AUX output terminals.
Press the key labeled with the amplifier symbol to enable or disable output; an annunciator on the key indicates when the 5725A is being used.