709/709H/710 Precision Loop Calibrator Users Manual
The Fluke 709/709H/710 Precision Loop Calibrator is designed for the installation, calibration, and troubleshooting of field transmitters, valves, and other control system components. The device supports primary functions such as sourcing and measuring mA signals in the 0 to 24 mA range and producing up to 24 V dc loop power. The 709H model includes HART communication functionality, while the 710 adds specific test functions for HART smart control valves.
| Current measurement, sourcing range | 0 mA to 24 mA |
|---|---|
| Voltage measurement range | 0 V dc to 30 V dc |
| mA Resolution | 1 μA |
| Voltage Resolution | 1 mV |
| Accuracy (all ranges) | 0.01 % ±2 LSD |
| Operating Temperature Range | -10 °C to 50 °C |
| Storage Temperature Range | -20 °C to 60 °C |
| Display Type | 128 x 64 pixels, LCD Graphic with backlight |
| Battery Life (Alkaline) | ≥40 hours continuous use (measure mode) |
The device can be used for installation, calibration, and troubleshooting of field transmitters, valves, and other control system components.
The accuracy for all ranges is 0.01 % ±2 LSD at 23 °C ±5 °C.