This document is the warranty manual for the 772/773 Milliamp Process Clamp Meter Calibration Manual product, with a warranty period of 3 years (1 year for cable and clamp)
This manual provides instructions for qualified personnel to perform verification and adjustment procedures for the Fluke 772/773 Milliamp Process Clamp Meter. It details safety precautions, electrical and mechanical specifications, and basic maintenance tasks such as cleaning and battery replacement.
The document also outlines specific calibration and verification procedures, including accuracy tests for mA DC clamp, mA DC, and Volts DC measurements, as well as source accuracy tests. It includes step-by-step instructions for calibration adjustments and lists replaceable parts.
| DC Current Measurement With Jaw Ranges | 0 mA to 20.99 mA, 21 mA to 100 mA |
|---|---|
| In Circuit Range | 0 mA to 24 mA |
| Maximum Voltage | 50 V DC |
| Voltage Measurement (773) | 0-30 V |
| Batteries | 4 1.5 V, Alkaline, IEC LR6 |
The data in this manual is for qualified personnel only.
Calibration adjustment is required after a repair or if the Meter fails a performance test.
Calibration adjustment is required if the Meter fails a performance test.