This is a Fluke 365 Detachable Jaw True-rms Clamp Meter Calibration Manual Limited Warranty And Limitation Of Liability
This document serves as a calibration manual for the Fluke 365 Detachable Jaw True-rms Clamp Meter. It details the safety precautions required before beginning the process and lists the necessary equipment, including specific calibrator outputs and power supplies.
The core of the manual outlines a step-by-step adjustment procedure. This involves applying specific voltage, resistance, and current values from a calibrator to the meter's input terminals, waiting for the reading to stabilize, and confirming the steps to complete the calibration.
| AC Current Range | 200.0 A |
|---|---|
| AC Voltage Range | 600.0 V |
| DC Voltage Range | 600.0 V |
| Resistance Range | 600.0 Ω /6000 Ω |
| Size | 225 x 65 x 46 mm |
| Weight | 275 g |
| Batteries | 2 AA, NEDA 15A, IEC LR6 |
Read the Safety Information section before using the product to prevent electrical shock, fire, or personal injury.
A calibrator capable of outputting specific voltages (0V to 600V), resistances (0Ω to 6000Ω), and currents (0A to 10A) at various frequencies.
You must wait 4 seconds for the reading to stabilize before pushing the confirm button.