User Guide
The document serves as a comprehensive user manual for the ResMed S8 Series II family of Positive Airway Pressure (PAP) devices, specifically the S8 AutoSet™ II and S8 Elite™ II models. It details indications for use, contraindications, and potential adverse effects associated with the therapy. The guide explains how to set up the system, including connecting power and air tubing, selecting compatible masks, and utilizing specific features like the mask-fitting function and SmartStart™ automatic start.
| Operating pressure range | 4 to 20 cm H2O |
|---|---|
| Dimensions (H x W x D) | 4.6" x 6.5" x 5.7" (112 mm x 164 mm x 145 mm) |
| Weight | 3.1 lb (1.4 kg) |
| Operating Temperature | +5°C to +36°C (+41°F to +95°F) |
| Input range (with HumidAire 3i) | 100–240V , 50/60Hz; 110V , 400Hz |
| Operating Altitude | Sea level to 8500' (2591 m) |
The S8 AutoSet II is indicated for the treatment of obstructive sleep apnea (OSA) in adult patients and has two treatment modes: AutoSet and fixed-pressure CPAP.
Positive airway pressure therapy may be contraindicated in patients with severe bullous lung disease, pneumothorax, pathologically low blood pressure, dehydration, or cerebrospinal fluid leak.
Potential side effects include drying of the nose, mouth, or throat; nosebleed; bloating; ear or sinus discomfort; eye irritation; and skin rashes.
This feature delivers air pressure for a three-minute period prior to starting treatment to help check and adjust mask fit, indicated by asterisks on the LCD screen.