ResMed Mask System Disinfection and Cleaning Guide

This guide provides disinfection and cleaning instructions for ResMed mask systems (full face, nasal pillows, and nasal masks) for reuse between patients in…


ResMed Mask System Disinfection and Cleaning Guide - cover page
Brand
ResMed
Category
Other
Document type
Service Manual
Language
Chinese
Pages
4
File format
PDF
File size
216 KB
Published
20 October, 2026
Updated
20 October, 2026
MD5 checksum
DD452856C7C48025A9BF6FEC0A8DE931

About this manual

This document serves as a disinfection guide for ResMed mask systems, including full face masks, nasal masks, and nasal pillows, intended for reuse between multiple patients in clinical environments such as sleep laboratories, clinics, or hospitals. It outlines specific procedures for manual hot water disinfection and automated cleaning/disinfection cycles using a washer-disinfector. The guide specifies the maximum number of validated reuse cycles for various components and lists compatible cleaning agents, such as Alconox and neodisher MediClean forte. It also details the necessary disassembly steps before cleaning and inspection criteria for damaged components. This manual is written in Chinese.

What's inside

  • Disinfection Guide
  • Manual Disinfection Process
  • Washer-Disinfector Process
  • Validated Reuse Cycles
  • Compatible Cleaning Agents
  • Inspection and Drying
  • Reassembly and Storage
  • Safety Warnings

Specifications

Compatible Cleaning Agent 1Alconox 1% concentration (10g/L)
Compatible Cleaning Agent 2neodisher MediClean forte
Manual Wash Temp Range30-55 degrees Celsius
Manual Wash Soak Time5-10 minutes
Manual Rinse Temp Range20-60 degrees Celsius
Manual Rinse Time (Components)30-60 seconds
Manual Rinse Time (Headband)1 minute
Manual Disinfection Temp/Time Option 170 degrees Celsius, 100 minutes
Manual Disinfection Temp/Time Option 275 degrees Celsius, 30 minutes
Manual Disinfection Temp/Time Option 380 degrees Celsius, 10 minutes
Manual Disinfection Temp/Time Option 490-93 degrees Celsius, 1-10 minutes

Frequently asked questions

What is the maximum validated reuse cycle for the Ultra Mirage full face mask valve?

The maximum validated reuse cycle for the Ultra Mirage valve is 15 times.

Can autoclaving or ethylene oxide gas sterilization be used on these masks?

No, the guide explicitly states not to use autoclaving or ethylene oxide gas sterilization for disinfection.

What water quality is required for cleaning and disinfection steps?

All cleaning, rinsing, and disinfection steps must use drinking quality water.