This guide provides disinfection and cleaning instructions for ResMed mask systems (full face, nasal pillows, and nasal masks) for reuse between patients in…
This document serves as a disinfection guide for ResMed mask systems, including full face masks, nasal masks, and nasal pillows, intended for reuse between multiple patients in clinical environments such as sleep laboratories, clinics, or hospitals. It outlines specific procedures for manual hot water disinfection and automated cleaning/disinfection cycles using a washer-disinfector. The guide specifies the maximum number of validated reuse cycles for various components and lists compatible cleaning agents, such as Alconox and neodisher MediClean forte. It also details the necessary disassembly steps before cleaning and inspection criteria for damaged components. This manual is written in Chinese.
| Compatible Cleaning Agent 1 | Alconox 1% concentration (10g/L) |
|---|---|
| Compatible Cleaning Agent 2 | neodisher MediClean forte |
| Manual Wash Temp Range | 30-55 degrees Celsius |
| Manual Wash Soak Time | 5-10 minutes |
| Manual Rinse Temp Range | 20-60 degrees Celsius |
| Manual Rinse Time (Components) | 30-60 seconds |
| Manual Rinse Time (Headband) | 1 minute |
| Manual Disinfection Temp/Time Option 1 | 70 degrees Celsius, 100 minutes |
| Manual Disinfection Temp/Time Option 2 | 75 degrees Celsius, 30 minutes |
| Manual Disinfection Temp/Time Option 3 | 80 degrees Celsius, 10 minutes |
| Manual Disinfection Temp/Time Option 4 | 90-93 degrees Celsius, 1-10 minutes |
The maximum validated reuse cycle for the Ultra Mirage valve is 15 times.
No, the guide explicitly states not to use autoclaving or ethylene oxide gas sterilization for disinfection.
All cleaning, rinsing, and disinfection steps must use drinking quality water.