This document is the user manual of the EchoPod® Ultrasonic Level Transmitter. It introduces the features of the product, including new features, specifications, dimensions, reflective technology, and configuration methods.
This manual provides instructions for the Flowline EchoPod® Ultrasonic Level Transmitter, model MN310140. It covers the reflective ultrasonic technology used for measuring liquid levels, including setup using either the WebCal® software or the internal display module. The document details the installation process, wiring diagrams, and configuration steps for both UG06 and UG12 series models.
| Supply Voltage | 14-28 Vdc |
|---|---|
| Signal Output | 4 to 20 mA current loop |
| Temperature Range | -40 to 80 °C |
| Process Mount UG06 | 2" NPT or 2" G |
| Process Mount UG12 | 3" NPT or 3" G |
| Range UG06 | 6.0m (19.69 feet) |
| Accuracy | 0.2% of Range in air at 20 °C |
The EchoPod can be configured using either the free WebCal® software and USB® Fob or with the internal display module.
The maximum supply voltage is 28 Vdc.
No, the EchoPod should not be used within classified hazardous environments.