Flowline MN300820 Echotouch Ultrasonic Level Transmitter User Manual

Echotouch Ultrasonic Level Transmitter is a new reflective ultrasonic level transmitter with simple configuration and wider temperature range.


Flowline MN300820 Echotouch Ultrasonic Level Transmitter User Manual - cover page
Brand
FlowLine
Category
Industrial Robot
Document type
User Manual
Language
English
Pages
64
File format
PDF
File size
2.9 MB
Published
12 October, 2023
Updated
12 October, 2023
MD5 checksum
245FAED0CB4D92303883262C057DE590

About this manual

This manual provides instructions for the Flowline MN300820 Echotouch Ultrasonic Level Transmitter, available in US01, US03, and US06 series. It details the reflective ultrasonic technology used to measure liquid levels by orienting the transducer vertically to prevent condensation interference. The device measures the time of flight of sound waves reflected from the process medium to determine distance.

What's inside

  • Introduction
  • Specifications
  • Dimensions
  • Intrinsically Safe Control Drawings
  • Safety Precautions
  • Getting Started
  • Setup Overview
  • WebCal Configuration
  • Display Configuration
  • Wiring

Specifications

Supply Voltage14-28 Vdc
Signal Output4 to 20 mA current loop
Range (US06)6 meters (19.69 feet)
Range (US03)3 meters (9.84 feet)
Range (US01)1.5 meters (4.92 feet)
Accuracy (US06, US03)0.2% of Range in air at 20 °C
Accuracy (US01)0.125" (3mm)
Resolution0.125" (3mm)
Temperature Range (F)-40° to 176°
Enclosure Rating (US06)TYPE 4X, IP66
Enclosure Rating (US01, US03)TYPE 6P, IP68

Frequently asked questions

How does the EchoTouch Ultrasonic Level Transmitter handle condensation?

It uses Reflective Technology by orienting the transducer vertically so condensation runs off the smooth surface without affecting signal performance.

What is the maximum supply voltage for the transmitter?

The maximum supply voltage is 28 Vdc.

Can the unit be configured using a handheld device?

Yes, the US06 series supports HART signal when a display is used, allowing configuration with a HART handheld communicator.