HTCK-XB is an OEM temperature and humidity transmitter designed by ROTRONIC for high and low temperature industrial applications. It features high accuracy, wide temperature range, and digital calibration.
The ROTRONIC HTCK-XB is an OEM high-low temperature industrial temperature and humidity transmitter designed using digital technology. It features a split design between the sensor and transmission circuitry, offering high precision (±1.5%RH, ±0.3°C), wide operating temperatures (-75°C to +200°C), and software-based calibration for improved stability and accuracy. This manual is written in Chinese.
| Measurement Accuracy (Humidity) | ±1.5%RH |
|---|---|
| Measurement Accuracy (Temperature) | ±0.3°C (@23 ± 5 ℃ ) |
| Operating Temperature Range | -75 … +200 ℃ |
| Power Supply Voltage | 15 … 40VDC (Recommended: 24VDC) |
| Current Consumption | Max. 50mA |
| Output Load | Max. 500 Ω |
The typical response time (T63) is 10 seconds under a 1m/s airflow.
The dedicated AC3006 communication cable must be used; standard Mini USB cables will cause irreversible damage.
No, cutting or reconnecting the 2-meter connection wire between the probe and circuit will cause measurement inaccuracy due to parasitic capacitance.