2511 DC Low Resistance Tester Operation Manual

This document is the operation manual for the 2511 DC low resistance tester, including precautions for the instrument's power-on preheating, zeroing, calibration, etc., as well as the instrument's functions and usage conditions.


2511 DC Low Resistance Tester Operation Manual - cover page
Brand
Unknown
Category
Other Electronics
Document type
User Manual
Language
Chinese
Pages
8
File format
PDF
File size
265 KB
Published
14 December, 2022
Updated
28 September, 2023
MD5 checksum
56E79A5D5D4F4D2CEDC93CC900BF4B9A

About this manual

The 2511 DC Low Resistance Tester is an intelligent instrument designed for high-precision four-terminal measurement of low resistance values, ranging from 1uΩ to 200.00kΩ. It features automatic range selection, sorting functions with LED indicators (HIGH, PASS, LOW), and a test speed of 10 to 20 times per second. The device utilizes a 5-digit LED display and includes a built-in EEPROM for storing calibration parameters.

The manual details the operational procedures, including the necessity of a 10 to 30-minute warm-up period before testing to ensure internal circuit stability. It outlines the use of five-terminal connections (HD, HS, LD, LS, GND) for accurate measurements and provides specific steps for zeroing the instrument and calibrating it using standard resistors. Safety precautions emphasize avoiding live testing and proper grounding to prevent electrical hazards. This manual is written in Chinese.

What's inside

  • Overview
  • Functions
  • Operating Conditions
  • Technical Specifications
  • Front Panel Description
  • Zeroing and Calibration
  • Test Port Description
  • Accuracy Verification
  • Unboxing
  • Warranty

Specifications

Measurement Range1uΩ~200.00kΩ
Test Speed10~20 times/second
Power SupplyAC220V±10%, 50Hz±5%
Operating Temperature20~26℃ (accuracy), 10~35℃ (operation)
Dimensions300×110×280mm
WeightApprox. 2.5kg
Display5-digit LED

Frequently asked questions

How long must the instrument warm up before testing?

The instrument must warm up for at least 10 minutes, or 30 minutes according to the unboxing section, before measurement to allow internal circuits to stabilize.

What is the warranty period for this instrument?

The warranty period is 18 months from the date of shipment by the company or the operating department.

Can I replace internal chips if the instrument malfunctions?

No, the manual strictly prohibits replacing any internal chips as they contain stored calibration parameters that would make measurements incorrect.