Extech Models 601100/601500/60120B pH Electrodes Instruction Sheet

601XXX V1.3 11/01 pH electrode instruction sheet introduces the specifications of the 601100, 601500 and 60120B models, including dimensions, construction material, measuring range, temp. range, reference junction, reference type, isopotential point, precision, connector, cable length and storage container


Extech Models 601100/601500/60120B pH Electrodes Instruction Sheet - cover page
Brand
EXTECH
Category
Other Electronics
Document type
User Manual
Language
English
Pages
1
File format
PDF
File size
125 KB
Published
01 August, 2012
Updated
01 October, 2023
Model
60120B
MD5 checksum
B613F9613BB06452DE13141AD28CCC74

About this manual

This document serves as an instruction sheet for three specific Extech pH electrode models: the flat surface combination model (601100), the mini combination model with detachable bulb guard (601500), and the general-purpose combination model (60120B). It provides detailed specifications including dimensions, measuring range of 0 to 14.0 pH units, and temperature ranges from 0-80°C.

What's inside

  • Electrode Specifications
  • General Electrode Procedure
  • Before Measurement
  • pH Calibration and Measurement
  • Two-Buffer Calibration
  • Single-Buffer Calibration
  • pH Measurement
  • Electrode Storage and Maintenance
  • Cleaning Procedures
  • Warranty

Specifications

Measuring range0 to 14.0 pH units
Temp. range0-80 o C
Reference typeAg/AgCl (silver silver chloride)
Precision0.02 pH units
ConnectorBNC
Cable length1 meter (39")

Frequently asked questions

What is the recommended storage solution for electrodes 601100 and 601500?

Store in a soaking bottle with 22.4g KCl dissolved in 100ml pH4 buffer solution or 25g KCl dissolved in 100ml pure water.

How long does the warranty last for these instruments?

Extech warrants the instrument to be free of defects for one year from date of shipment, with a six-month limited warranty on sensors and cables.

What is the procedure if an electrode has been allowed to dry?

Soak the electrode in a pH 4 buffer solution for at least 20 minutes, which may require up to 24 hours depending on dryness.

Download PDF (125 KB)