Rosemount Analytical Hx348 Pressurized Mounting Assembly Instruction Sheet


Rosemount Analytical Hx348 Pressurized Mounting Assembly Instruction Sheet - cover page
Brand
Emerson
Category
Industrial Supply
Document type
Installation Guide
Language
English
Pages
2
File format
PDF
Published
10 July, 2012
Updated
29 September, 2023
MD5 checksum
E08483916FD167569D7237C4441195AF

About this manual

The document is an instruction sheet (PN 51A-PRES/rev.A, February 2004) for the Pressurized Mounting Assembly designed for use with Model Hx348 or sensors with a 12-mm lower sensor shaft and a pressurized or disposable reference design. It lists wetted parts of stainless steel and Viton, a maximum temperature of 130°C, and a maximum pressure of 72 psig.

Installation steps cover mounting the sensor, installing the cable, and mounting the assembly into a G 1-1/4 in. weld-in socket. Cleaning and maintenance notes describe cleaning fluids and replacement of Viton o-rings using the Pressurizable Assembly Servicing Kit (PN 9160485).

What's inside

  • Specifications
  • Installation Notes
  • Parts Locator
  • Mounting the Sensor into the Pressurized Mounting Assembly
  • To install the cable
  • Mounting the Mounting Assembly into the Process
  • Cleaning and Maintenance
  • Replacing the o-rings

Specifications

Wetted PartsStainless Steel (type AISI-ASTM 316 Ti; B.S. 320 $17) and Viton
Maximum Temperature130°C (266°F)
Maximum Pressure72 psig (600 kPa, 6 bar)
Process ConnectionG 1-1/4 in. weld-in socket
Sensor Shaft Diameter12 mm
Assembly Pressure Above Tank or Fermenter14 psi (100 kPa, 1 bar)

Frequently asked questions

What is the maximum pressure for the assembly?

The pressure in the assembly should not exceed 72 psig (600 kPa, 6 bar).

Which sensors can be used with the assembly?

It is designed for use with Model Hx348 or sensors with a lower sensor shaft that has a 12-mm diameter and a pressurized or disposable reference design.

How should the assembly be cleaned?

It can be cleaned with general-purpose cleaning fluids, and scouring agents that contain hard particles should never be used.

What kit is used to replace the o-rings?

Viton o-rings can be replaced with o-rings from the Pressurizable Assembly Servicing Kit (PN 9160485).