HANNA 981402C On-line pH Monitor for CO2 Dosing Control Instruction Manual

This manual introduces the use and precautions of HI981402C pH online monitor.


HANNA 981402C On-line pH Monitor for CO2 Dosing Control Instruction Manual - cover page
Brand
Hanna Instruments
Category
Other
Document type
User Manual
Language
English
Pages
2
File format
PDF
File size
616 KB
Published
10 July, 2012
Updated
30 September, 2023
MD5 checksum
44426B563A76A5D51E602CB0D920627C

About this manual

The Hanna Instruments HI 981402C is an accurate and safe instrument designed to dose CO2 in planted aquariums based on pH measurement. It operates using 12 Vdc power to prevent electric shock risks and is ideal for controlling CO2 levels and flow by monitoring pH values. When the measured pH exceeds a user-selectable setpoint, the instrument activates a magnet valve to stop dosing.

What's inside

  • Warranty
  • Functional Description
  • Specifications
  • pH Electrode Connection
  • Taking Measurements
  • Calibration
  • Adjusting the Setpoint
  • Accessories

Specifications

Range0.0 to 14.0 pH
Resolution0.1 pH
Setpoint Range5.0 to 8.0 pH
Hysteresis0.1 pH
Output12 Vdc/500 mA
Casing ProtectionIP54
Dimensions85 x 94 x 33 mm
Weight160 g

Frequently asked questions

What is the factory setpoint value?

The factory setting for the setpoint is 6.1 pH.

How often should the instrument be calibrated?

It is recommended to calibrate the instrument at least once a month.

What happens when the pH exceeds the setpoint?

The red alarm LED blinks and the power contact output activates to drive an external magnet valve.