Hanna Instruments HI 9214 Stick pH Temperature Meter Instruction Manual

HI 9214 Stick pH & Temperature Meter is a product of Hanna Instruments. The instruction manual introduces the use of the product and warranty policy


Hanna Instruments HI 9214 Stick pH Temperature Meter Instruction Manual - cover page
Brand
Hanna Instruments
Category
Other Electronics
Document type
User Manual
Language
English
Pages
2
File format
PDF
File size
474 KB
Published
10 July, 2012
Updated
30 September, 2023
MD5 checksum
230EACB5D2E36182DCA13EE92DD4A717

About this manual

The HI 9214 is a compact, stick-type, microprocessor-based pH and temperature meter designed for easy portability. It features an interchangeable plastic-body electrode with a built-in temperature sensor for automatic temperature compensation. The device includes three memorized buffer values (pH 4.01, 7.01, and 10.01) and automatic buffer recognition to simplify calibration for non-technical personnel.

The manual covers preliminary examination, functional descriptions of the controls, and detailed operational guides for taking measurements. It also explains how to replace the battery, perform single or two-point pH calibrations, and properly store the electrode. Safety recommendations and warranty information are also included.

What's inside

  • Warranty
  • Preliminary Examination
  • General Description
  • Functional Description
  • Specifications
  • Operational Guide
  • Battery Replacement
  • pH Calibration
  • Safety Recommendations

Specifications

Range-2.00 to 16.00 pH / -5.0 to 70.0 °C
Resolution0.01 pH / 0.1 °C
Accuracy±0.02 pH / ±0.5 °C
Battery Type1x 9V, alkaline
Battery Life700 hours of continuous use
Auto-offAfter 8 minutes of non-use
Dimensions185 x 42 x 26 mm
Weight160 g

Frequently asked questions

What is the warranty period for the meter?

All Hanna Instruments meters are warranted for two years against defects in workmanship and materials.

What happens when the battery is low?

The low battery symbol lights up on the LCD, and the meter displays the 'Eb' message at startup.

Can the meter recognize buffer values automatically?

Yes, the meter has three memorized buffer values and automatic buffer recognition to avoid errors during calibration.

What should be used to store the electrode?

The electrode should be stored with a few drops of storage solution (HI 70300) or pH 7 solution in the protective cap; distilled water must not be used.