Hanna Instruments HI 38001 Sulfate Test Kit Instruction Manual

HI 38001 Sulfate Low and High Range Test Kit is used for measuring the concentration of sulfate in water. The kit contains sulfate reagent, sulfate solution, plastic vessel, syringes etc. It is easy to use.


Hanna Instruments HI 38001 Sulfate Test Kit Instruction Manual - cover page
Brand
Hanna Instruments
Category
Other
Document type
User Manual
Language
English
Pages
1
File format
PDF
File size
58 KB
Published
09 July, 2012
Updated
30 September, 2023
MD5 checksum
9D2B8C6695DE0BC7511B27893959C408

About this manual

This document is the instruction manual for the Hanna Instruments HI 38001 Sulfate Low and High Range Test Kit. It outlines the contents of the kit, including reagents and vessels, and explains the chemical reaction used for sulfate determination via titration with Sulfonazo II.

The manual provides step-by-step instructions for performing tests in two ranges: Low Range (100 to 1000 mg/L) and High Range (1000 to 10000 mg/L). It details the procedure for preparing samples, adding reagents, and calculating sulfate concentration based on the volume of titration solution used. Safety warnings advise reading the Health and Safety Data Sheet before handling chemicals.

What's inside

  • Contents
  • Significance and Use
  • Chemical Reaction
  • Low Range Instructions
  • High Range Instructions
  • Calculations
  • Health and Safety

Specifications

Range Low100 to 1000 mg/L as Sulfate
Range High1000 to 10000 mg/L as Sulfate
Smallest Increment Low10 mg/L Sulfate
Smallest Increment High100 mg/L Sulfate
Analysis MethodTitration
Number of Tests200
Case Dimensions235x176x115mm

Frequently asked questions

What is the analysis method for the HI 38001 kit?

The analysis method is titration, specifically a modification of the Determination of Sulfate by Sulfonazo II.

How do you calculate the sulfate concentration for the Low Range?

Read the milliliters of titration solution from the syringe scale and multiply by 1000 to obtain mg/L (ppm).

How do you calculate the sulfate concentration for the High Range?

Read the milliliters of titration solution from the syringe scale and multiply by 10000 to obtain mg/L (ppm).

What indicates the end point of the titration?

The end point is indicated by the change in color of the solution from violet to blue.