HI 38001 Sulfate Low and High Range Test Kit is used for measuring the concentration of sulfate in water. The kit contains sulfate reagent, sulfate solution, plastic vessel, syringes etc. It is easy to use.
This document is the instruction manual for the Hanna Instruments HI 38001 Sulfate Low and High Range Test Kit. It outlines the contents of the kit, including reagents and vessels, and explains the chemical reaction used for sulfate determination via titration with Sulfonazo II.
The manual provides step-by-step instructions for performing tests in two ranges: Low Range (100 to 1000 mg/L) and High Range (1000 to 10000 mg/L). It details the procedure for preparing samples, adding reagents, and calculating sulfate concentration based on the volume of titration solution used. Safety warnings advise reading the Health and Safety Data Sheet before handling chemicals.
| Range Low | 100 to 1000 mg/L as Sulfate |
|---|---|
| Range High | 1000 to 10000 mg/L as Sulfate |
| Smallest Increment Low | 10 mg/L Sulfate |
| Smallest Increment High | 100 mg/L Sulfate |
| Analysis Method | Titration |
| Number of Tests | 200 |
| Case Dimensions | 235x176x115mm |
The analysis method is titration, specifically a modification of the Determination of Sulfate by Sulfonazo II.
Read the milliliters of titration solution from the syringe scale and multiply by 1000 to obtain mg/L (ppm).
Read the milliliters of titration solution from the syringe scale and multiply by 10000 to obtain mg/L (ppm).
The end point is indicated by the change in color of the solution from violet to blue.