This instruction manual introduces the usage of Hanna iron high range test kit. The test kit can detect the iron content in water, and the detection range is 1-50 mg/L (ppm).
The Hanna HI 38041 is an Iron High Range Test Kit designed to determine total iron levels in water. The kit includes reagents, glass vials, pipettes, and a checker disc for visual comparison of results.
To perform the test, users fill vials with sample water and deionized water, add the HI 38041-0 reagent, wait three minutes for the reaction to turn orange, and then compare the color intensity against the provided checker disc. The kit measures iron concentrations ranging from 0 to 10.0 mg/L (ppm) with a smallest increment of 0.2 mg/L.
| Range | 0 to 10.0 mg/L (ppm) as Fe |
|---|---|
| Smallest Increment | 0.2 mg/L (ppm) Fe |
| Analysis Method | Colorimetric |
| Sample Size | 5 mL |
| Number of Tests | 100 |
Read the value in the result window and multiply it by 2 to obtain mg/L (or ppm) of Iron.
Hold the checker disc so that a light source illuminates the samples from the back of the windows.
Molybdate and Molybdenum above 50 ppm, calcium above 10000 ppm, magnesium above 100000 ppm, and chloride above 185000 ppm.