This document describes the Operational Qualification/Performance Verification (OQ/PV) procedures of Agilent Technologies for the 6890 series gas chromatograph (GC) with headspace sampler. It explains how to document the correct functioning of the entire system according to established specifications. The OQ/PV procedures use internal tests and analytical methods with certified samples. Typically, the entire OQ/PV procedure can be completed in one day, minimizing downtime and maintaining high productivity. However, the limits described in the document and the online Help system of ChemStation apply to new or repaired instruments according to Agilent specifications. Instruments that have been in daily use for a long time may not comply with Agilent specifications. Acceptance criteria may vary depending on analytical requirements and should be agreed upon with Agilent support services.
This document outlines the Operational Qualification (OQ) and Performance Verification (PV) procedures for Agilent Technologies Gas Chromatographs (GC) Series 6890 and 6850 equipped with a head space sampler. These procedures allow for documenting the correct functioning of the entire system according to established specifications using internal tests and analytical methods with certified samples. The entire OQ/PV procedure can typically be completed in one day to minimize downtime and maintain high productivity. Tests can be performed selectively and in any order, though following the prescribed sequence minimizes the time required to complete the procedure and the stabilization time. This manual is written in Italian.
| Supported Models | GC Agilent 6890 e 6850 |
|---|---|
| Verification Standard | ASTM E 685-93 |
Yes, the entire procedure can typically be completed in one day to minimize downtime.
Injector maintenance is recommended before the procedure.
No, the procedure does not include preventive maintenance or qualification of additional revolvers and injectors not specified in the configuration.