Agilent 6890/6850 GC Head Space Sampler OQ/PV Procedure

This document describes the Operational Qualification/Performance Verification (OQ/PV) procedures of Agilent Technologies for the 6890 series gas chromatograph (GC) with headspace sampler. It explains how to document the correct functioning of the entire system according to established specifications. The OQ/PV procedures use internal tests and analytical methods with certified samples. Typically, the entire OQ/PV procedure can be completed in one day, minimizing downtime and maintaining high productivity. However, the limits described in the document and the online Help system of ChemStation apply to new or repaired instruments according to Agilent specifications. Instruments that have been in daily use for a long time may not comply with Agilent specifications. Acceptance criteria may vary depending on analytical requirements and should be agreed upon with Agilent support services.


Agilent 6890/6850 GC Head Space Sampler OQ/PV Procedure - cover page
Brand
Agilent
Category
Other Electronics
Document type
Service Manual
Language
Italian
Pages
2
File format
PDF
File size
38 KB
Published
06 July, 2012
Updated
28 September, 2023
MD5 checksum
02E56F276B97F812515FC70CB2EED44C

About this manual

This document outlines the Operational Qualification (OQ) and Performance Verification (PV) procedures for Agilent Technologies Gas Chromatographs (GC) Series 6890 and 6850 equipped with a head space sampler. These procedures allow for documenting the correct functioning of the entire system according to established specifications using internal tests and analytical methods with certified samples. The entire OQ/PV procedure can typically be completed in one day to minimize downtime and maintain high productivity. Tests can be performed selectively and in any order, though following the prescribed sequence minimizes the time required to complete the procedure and the stabilization time. This manual is written in Italian.

What's inside

  • Requisiti OQ/PV
  • Hardware supportato
  • Procedure di prova
  • Vantaggi del servizio OQ/PV
  • Documentazione
  • Certificato OQ/PV

Specifications

Supported ModelsGC Agilent 6890 e 6850
Verification StandardASTM E 685-93

Frequently asked questions

Can the OQ/PV procedure be completed in one day?

Yes, the entire procedure can typically be completed in one day to minimize downtime.

What is the recommended maintenance before OQ/PV?

Injector maintenance is recommended before the procedure.

Does the procedure cover additional injectors or detectors?

No, the procedure does not include preventive maintenance or qualification of additional revolvers and injectors not specified in the configuration.