This document serves as a selection guide for Agilent products used in bioseparation. It provides strategies for developing reversed-phase HPLC methods for proteins and peptides, detailing column choices based on molecular weight and pH conditions.
The guide covers various column technologies including ZORBAX StableBond, Poroshell 300SB, and Bio-Monolith columns. It includes specific recommendations for analytical separations, LC/MS applications, and high-sensitivity capillary chromatography. This manual is written in German.
| Pore size for Poroshell 300SB-C18, C8, C3 | 300A |
|---|---|
| Pore size for Poroshell 300Extend-C18 | 300A |
| Particle size for Poroshell columns | 5 µm |
The StableBond 300SB-C8 column is widely recommended as the first choice due to its excellent lifetime in TFA-containing solvents.
A water/acetonitrile gradient containing 0.1% TFA is commonly used to elute all relevant components.
The ZORBAX 300Extend-C18 column is an excellent choice for separations performed at a medium or high pH value.
The specifications list a particle size of 5 µm for the various Poroshell column models.