Agilent 1100 Series HPLC System Optimization Technical Note

This document describes how to achieve the highest performance of Agilent 1100 Series HPLC system


Agilent 1100 Series HPLC System Optimization Technical Note - cover page
Brand
Optimizing
Category
Other
Document type
Service Manual
Language
English
Pages
8
File format
PDF
File size
465 KB
Published
05 July, 2012
Updated
30 September, 2023
MD5 checksum
EA77ACED0FE103433397DF6C12CE2341

About this manual

This document provides a standard check-out procedure to ensure the Agilent 1100 Series HPLC system is free of defects before optimization. It details steps for checking for leaks, contamination, and baseline noise, as well as verifying retention time and area precision using specific parameters and standard samples.

The guide further explains how to optimize performance for specific issues such as low sensitivity, poor precision, low flow rates, and carry-over. It includes recommendations for detector settings, pump configurations, and autosampler adjustments to achieve the best results in normal phase and gradient chromatography.

What's inside

  • Standard Check-out Procedure
  • Checking for Leaks and Contamination
  • Checking for Baseline Noise
  • Checking Retention Time and Area Precision
  • Optimizing Sensitivity
  • Optimizing Retention Time and Precision
  • Optimizing for Low Flow Rates
  • Reducing Carry-over
  • Normal Phase Chromatography
  • System Shutdown Recommendations

Specifications

Variable Wavelength Detector (VWD) Noise Specification+ 0.75 x 10° AU at 254 nm
Diode Array Detector (DAD) Noise Specification+ 1x 10°AU at 254 and 750 nm
Retention Time Precision Requirement< 0.5 %
Area Precision Requirement< 2%

Frequently asked questions

What is the recommended flow rate for checking baseline noise?

The recommended flow rate is 1 ml/min.

What is the optimum reference wavelength for a sample with a maximum at 252 nm?

The optimum reference wavelength is 360 nm with a bandwidth of 100 nm.

What pump type is recommended for flow rates below 200 µl/min?

A binary pump with a degasser is recommended.

What should be done if the instrument will not be used for several days?

The system should be flushed with water for at least half an hour.