Agilent Prep-Sil Columns Datasheet

Agilent Prep-Sil is a preparative-grade liquid chromatography column produced by Agilent. It uses ultra-pure (Type B) silica gel with high specific surface area (400 m2/g) and 100Å pore size in two particle sizes - 5-microns and 10-microns. Agilent Prep-Sil columns have high sample loadability and produce symmetrical peaks even for strongly basic compounds. The columns are stable at pH 0.8-8. Agilent Prep-Sil is especially useful for the preparative purification of acidic, basic, and other highly polar compounds by normal-phase liquid chromatography. Columns are loaded to a stable, uniform bed density using a proprietary, high-pressure slurry-loading technique to give maximum column efficiency and maintain column bed stability.


Agilent Prep-Sil Columns Datasheet - cover page
Brand
Agilent
Document type
Datasheet
Language
English
Pages
2
File format
PDF
File size
85 KB
Published
04 July, 2012
Updated
30 September, 2023
MD5 checksum
BB22EA80E60A40F5986E15D094640E86

About this manual

This document provides a technical overview of Agilent Prep-Sil columns, which are packed with ultra-pure silica suitable for the preparative purification of acidic, basic, and highly polar compounds via normal-phase liquid chromatography. The columns feature high sample loadability and symmetrical peaks, utilizing proprietary slurry-loading techniques to ensure bed stability.

What's inside

  • General Description
  • Column Characteristics
  • Thermal Effects
  • Table 1 Typical Sample Capacities
  • Safety Considerations
  • Operational Guidelines
  • Preparative Strategies
  • Storage Recommendations

Specifications

Particle Size5-microns or 10-microns
Pore Size100A
Specific Surface Area400 m²/gram
pH Stability Range0.8-8
Maximum Operating Pressure840 bar (5000psi)

Frequently asked questions

What solvents are the Agilent Prep-Sil columns compatible with?

The columns are compatible with water and all common organic solvents.

What is the recommended pH range for operation?

Avoid use of this column below pH 0.8 or above pH 8.

How should large-diameter columns be operated to avoid thermal effects?

It is recommended that they be heated in a water bath at 30°C or higher.