Varian 212-LC Chromatography Pump Product Leaflet

This data sheet describes the features and performance of the Varian 212-LC liquid chromatography pump.


Varian 212-LC Chromatography Pump Product Leaflet - cover page
Brand
Agilent
Category
Other
Document type
Product Leaflet
Language
English
Pages
2
File format
PDF
File size
607 KB
Published
04 July, 2012
Updated
30 September, 2023
MD5 checksum
D5376AA26415513AD0ADDEECFF5AD74D

About this manual

The Varian 212-LC chromatography pump is a binary pumping system designed for stable and reliable analytical LC/MS applications. It is engineered to provide accurate, smooth flow with low delay volume, making it ideal for gradient and isocratic applications up to 2 mL/min. The pump is compatible with a range of column sizes, from 1 mm to 4.6 mm, and is suitable for environmental samples, small molecule pharmaceuticals, and industrial process samples.

This document outlines the pump's key features, including self-priming check valves and MS Workstation control for full integration. It provides technical specifications such as flow rate precision, gradient composition accuracy, and maximum pressure limits. The leaflet also includes experimental conditions demonstrating the pump's performance and ordering information.

What's inside

  • Ideal solvent delivery
  • Ease-of-use
  • Full line of chromatography system products
  • Experimental conditions
  • Gradient Flow rate
  • Environment
  • Flow precision
  • Flow accuracy
  • Gradient composition accuracy
  • Communications

Specifications

Gradient Flow rate100 pL/min to 2 mL/min
Operation temperature10 °C to 35 °C
Relative humidity5% to 95%
Flow precision+1.5%
Flow accuracy+1% or +1 pL/min whichever is greater
Gradient composition accuracy<+1% absolute
Gradient composition precision<+1% absolute
Pressure Pulsation<1.0% A P/P
Maximum pressure400 Bar (6000 psi)
Maximum gradient flow rate2.0 mL/min
CommunicationsRS232
ControlKeypad and MS Workstation 6.6 or higher
Dimensions260 mm (w) x 330 mm (h) x 360 mm (d)
Weight15 kg/ 32 Ibs

Frequently asked questions

What is the flow rate range of the Varian 212-LC pump?

The pump delivers a flow rate from 100 pL/min to 2 mL/min.

What is the maximum pressure the pump can handle?

The maximum pressure is 400 Bar (6000 psi).

How does the pump communicate with the MS Workstation?

The pump uses RS232 for communications.