Agilent GPC of poly(n-isopropylacrylamide) Smart Material with Universal Calibration Using the 390-LC Multi Detector Suite (MDS) Application Note

This application note introduces the GPC analysis method for poly(n-isopropylacrylamide) (PNIPAM) smart materials using the Agilent 390-MDS multi-detector suite and universal calibration method.


Agilent GPC of poly(n-isopropylacrylamide) Smart Material with Universal Calibration Using the 390-LC Multi Detector Suite (MDS) Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
611 KB
Published
12 June, 2012
Updated
01 October, 2023
Model
390-LC
MD5 checksum
A7D1A6F89CE36728B1F95FBC217A4A9D

About this manual

The document explains how 'smart materials' like Poly(n-isopropylacrylamide) (PNIPAM) hydrogels change physical properties in response to stimuli such as temperature. Accurate measurement of their molecular weight distribution is critical for controlling drug release and uptake properties.

Standard GPC using only concentration detectors provides results relative to calibration standards, which may not be chemically identical to the sample. This note details a method using the Agilent 390-MDS Multi Detector Suite with online viscometry and differential refractive index (DRI) detection. By employing universal calibration based on molecular size in solution rather than just chemical identity, this technique provides accurate absolute molecular weights even when standards differ from the sample polymer.

What's inside

  • Introduction
  • GPC of Poly(n-isopropylacrylamide)
  • Universal Calibration Technique
  • 390-MDS Multi Detector Suite
  • Methods and Materials
  • Results and Discussion
  • Conclusion

Specifications

Columns2x Agilent PLgel 5 pm MIXED-C, 300 x 7.5 mm
EluentTHF + 5% triethylamine
Flow Rate1.0 mL/min
Detector Train390-MDS incorporating Viscometer and DRI
Detector TempAll detectors set at 40 °C

Frequently asked questions

What is the benefit of using universal calibration for PNIPAM?

Universal calibration allows accurate calculation of molecular weights even when polymer standards are not chemically identical to the sample, as it separates based on size rather than molecular weight.

Which detectors are used in the 390-MDS for this application?

The setup uses a viscometer and a differential refractive index (DRI) detector within the 390-MDS.

How does conventional GPC differ from universal calibration?

Conventional GPC with a concentration detector gives only relative results based on standards, whereas viscometry employing the 390-MDS provides an accurate representation of molecular weights regardless of calibrants used.

Download PDF (611 KB)